Description
VCA100CV Handheld Veterinary Vital Sign Monitor
- Feature:
- * Smallest and exquisite handheld vital sign monitor
- * Low perfusion reach 0.5%,perfect anti-motion
- * Provide Pl(perfusion index 0.05% – 20%
- * Suitable all kinds of animals
- * 5 inch TFT color screen
- * Two measurement mode:spot, monitor
- * NIBP adopt SAWAA(Self-Adaptive-Waveform-Amplitude-Adjust) algorithm and SAFD (Self-Adaptive-Fast-Deflation) algorithm
- * Data storage function:support data review, data trend and chart
- * Monitor mode,support continuous record for 48 hours
- * Spot mode,100 user ID, each use can storage 200 sets data
- * Support audio and visual alarm,can set low and high limit values
- * Lithium ion battery inside
- * Optional charging dock
- Multi Parameters for Collocation Optional:
- NIBP+SPO2+TEMP
- NIBP+TEMP
- SPO2+TEMP
- SPO2+ETCO2
- SPO2+ ETCO2 + NIBP
Technical Data |
|
Screen |
3.5″ TFT |
Resolution |
320*480 |
Indicator |
Alarm light, power light |
Waveform |
SPO2 and wavevalue |
Alarm |
Sensor off, low power, etc |
Alarm type |
Audio and visual |
Animal |
Small or big |
Power supply |
Lithium ion battery Operation time: 8 hours for normal operation |
TEMP |
Operation: 0-45°C: Storage: -25-65°C |
Humidity |
Operation: 15% to 90% non condensing |
Storage |
0% to 95% non condensing |
Altitude |
Up to 5000 meters |
Dimension |
146mm*67mm*30mm |
Weight |
250g |
NIBP |
|
Support big or small animals |
|
NIBP measurement range |
Big Animal Small Animal SYS 40-270mmHg 40-270mmHg MEAN 20-230mmHg 20 230mmHg DIA 10-210mmHg 10-210mmHg |
Static pressure range |
0-300mmHg |
NIBP measurement accuracy |
3mmHg |
Over pressure protection |
Software: 300mmHg Hardware: 320-330mmhg |
SPO2 |
|
Measurement range |
SpO2: 0-100% PR: 0-300bpm Perfusion Index: 0.05%-20% |
Accuracy range |
SpO2: 70%-100% PR: 30-254bpm Perfusion Index: 0.05%-20% |
Accuracy |
SpO2: ±2(70%-100%) Motion: ±3 PR: ±3bpm Motion: ±5bpm |
TEMP |
|
Measurement range |
25-45°C |
Resolution |
0.1°C |
Accuracy |
0.1°C |
CO2 |
|
Measurement range |
ETCO2: 0-150mmHg (0-19.79) (0-20kPa) RR: 3-150BPM |
Accuracy range |
ETCO2 Range Accuracy 0-40mmHg ±2mmHg 41-70mmHg ±5% of reading 71-100mmHg ±8% of reading 101-150mmHg ±10% of reading RR Accuracy ±1BPM |
Resolution: |
ETCO2: 0.1mmHg RR: 1BPM |
Setting compensation parameters |
Pressure range: 450-850mmHg Oxygen range: 0-100% Equilibrium gas: N2O, He |