VCA15M+ Veterinary Monitor

$0.00

SKU: VCA15M+ Category:

Description

VCA15M+ Veterinary Monitor

  • Feature:
  •  
  • * Professional design for vet: Cats (Small), Dogs (Middle), Horses (Large)
  • * 15″ color TFT display, high resolution: 1024x768dpi
  • * Standard 5-para: ECG, NIBP, TEMP, RESP, SPO2(Pr)                 
  • * Color: Grey; Light blue
  • * Special NIBP board for vet, alarm range, and software
  • * Special accessories for vet:
  • 3/5 lead ECG cable with vet limbs clip
  • Sp02 sensor with tongue and ear clip
  • NIBP cuff (cuff 1#, 2#, 3#, 4#, 5#)
  • Rectum TEMP sensor
  • * Separated para module: ECG module, NIBP module, SpO2 module
  • * Full touch-screen enables intuitive operation +navigation knob
  • * Sp02 Pitch Tone Variation;
  • * Multi- display: Standard, Large font, Trend Coexis, OxyCRG dynamic
  • * Standby mode for Energy saving and suspend monitoring;
  • * Screen brightness adjustable
  • * Capture dynamic waveforms
  • * Built-in rechargeable lithium battery, and battery volume checking
  • * AC power checking
  • * Efficient resistance to interference of defibrillator and HF knife
  • * Time delay switch-off function keeps the monitor data
  • * Complete and flexible mounting solution for transportation
  • * Optional/Comments: ETCO2, 2-IBP, Printer, Touchscreen, 3/12-lead ECG, Multi-gas(AG), BIS, SpO2, NIBP, Wall Mount, Trolley
  • * Package: 1PC/CTN: 45*24*43CM, G.W: 5.8KGS, 2PCS/CTN: 50*47*46CM, G.W: 12.7KGS

Technical Data

ECG

Lead selection: Standard 5 leads: I, II, III, AVR, AVL, AVF, V

ECG amplification width: 0.05Hz-100 Hz

S-T section monitoring: -2.0mV~2.0mV

Electrode shedding indication: Visual, audible and electric indicate

Scanning speed: 6.25, 12.5, 25, 50mm/SEC

Gain selection: x1/2, x1, x2, x4

ECG calibration: 1mV+5%

Protection Withstand 4000VAC/50Hz voltage isolation;

Alarm: The upper and lower limit of alarm value can be set, automatic memory, and alarm review

ECG cable: Standard: Animal specific 3-lead ECG cable

Heart Rate

Detection range of heart rate: 0-350bpm

Resolution: 1bpm

Accuracy: ±1%

SpO2

Display: Blood oxygen value, perfusion index, pulse histogram,

waveform, pulse value

SpO2 Range: 0%- 100% for large animals or small animals

Resolution: 1%

Accuracy: +2%(Undefined below 70%)

Pulse range: 0-300bpm

Pulse accuracy: ±1bpm

Standard sensors: Special probe for animal tongue and ear

Alarm value: The upper and lower limit of alarm value can be

set, automatic memory

Pulse Rate

Measurement range: 0~350

Alarm range: Upper limit: 80 to 350; Lower limit: 20 to 150

NIBP

Measurement method: Digital automatic oscillometric

Measurement parameters: Systolic pressure, Diastolic pressure,

Mean pressure, Pulse rate

Working mode: Manual, automatic, and continuous measurement

Unit: mmHg/KPa Optional

Interval time of automatic measurement: 0~250min adjustable

Overvoltage protection: Hardware and software protection

Measurement range: Large animals: 10mmHg~300mmHg

Small animals: 10mmHg~200mmHg

Cuff pressure range: 0~320mmHg

Standard cuff: 1#, 2#, 3#, 4#, 5# total 5PCS

Aeration time: Less than 6 seconds

Test vent time: Less than 15 seconds

TEMP

Measurement range: 0°C~50°C

Resolution: 0.1°C; Accuracy: ±0.1°C

Standard probe: Esophagus/rectum temp probe

Alarm: Upper and lower limit value can be set, automatic memory

RESP

Measurement method: Thoracic impedance

Respiration rate (RR):

A). Measurement range: 0–120bpm

B). Measurement accuracy: ±1bpm or 5%bp

Alarm: Upper and lower limit value can be set, automatic memory

2-IBP(Optional)

Number of channels: 2

Pressure sensor nominal sensitivity: 5uV/N/mmHg, error ±2%

Pressure sensor input impedance, output impedance: 3000 ~30002

Static pressure measurement range: -50~+300mmHg (Up to 350 mmHg)

Static pressure measurement error:

±1mmHg or ±1%, greater (Not including probe)

±4mmHg or ±4%, greater (Including probe)

Dynamic pressure measurement range: -50~+300mmHg

Dynamic pressure measurement error: ±4mmHg or ±4% of reading (greater)

Vessel types: ART, PA/ CVP/ RAP/ LAP/ ICP

Scope of stress test:

ART: 0~300mmHg

AP/RAP/LAP/ICP: -50~120mmHg

Resolution: 1mmHg

EtCO2(Optional)

Sampling Type: Mainstream/Side stream

Sampling Velocity:

Range: 50-150 ml/min

Accuracy: ±20%

Default value: 75m/min

Measurement principle: Non-spectral infrared, double channel, no moving parts(degassing pump)

 