Description
VCA3T Handheld Veterinary Vital Sign Monitor
- * The smallest and exquisite handheld vital signmonitor in the worldwide.
- * Accurate measurement in 0.05% Low perfusionIndex, the word leading production in the movement environment
- * Provide the parameter of PI( Perfusion Index05%-20%)
- * Accurate measurement in all kind of extremeenvironment measurement such as neonatal and shock patients
- * NIBP adoptSAWAA (Self-Adaptive-Waveform-Amplitude-Adjust) algorithm and SAFD (Self-Adaptive-Fast-Deflation) algorithm
- * Easy operation by5’’color TFT, resolution: 320*480
- * Up to 100 users ID switch
- * Multiple vital signs measurement parametersfor collocation optional
- NIBP, Spo2, Temp(VCA3T)
- SpO2 + NIBP + Main Stream EtCO2(VCA3E)
- * Two measurements mode: Spot, Monitoring
- * Data storage capabilities, support for historicaldata view and trend chart back to see
- In monitoring mode, can continuously record 48 hours of measurement data
- In spot mode, 100 users ID, each user can store 200 sets data
- * USB data upload
- * Support audio and visual alarm, support to setalarm upper and lower limit values
- * Lithium ion battery power supply
- * Optional Charging dock
Specification |
|
SpO2 |
Reusable SpO2 sensor |
NIBP |
NIBP extension cable NIBP Disposable Cuff (5 sizes) |
TEMP |
Reusable Temperature Probe |
Others
|
Charging Stock Power charger |
Technical Data |
|
NIBP |
|
The wider NIBP measurement range |
Big animal Small animal SYS 40 ~ 270mmHg 40 ~ 200mmHg MEAN 20 ~ 230mmHg 20 ~ 165mmHg DIA 10 ~ 210mmHg 10 ~ 150mmHg Pulse rate range: 40~240bpm |
NIBP measurement accuracy |
±3mmHg |
Pulse rate measurement accuracy |
<=2% (>=100 BPM) or 2 beats/min(<100 BPM) |
Independent software protection pressure |
Big animal: 300mmHg Small animal: 300mmHg |
Independent hardware protection pressure |
Big animal: 320-330mmHg Small animal: 320-330mmHg |
Oxygen Saturation (%Sp02) |
|
Measurement range |
* Measure range: SpO2: 0-100% PR: 0-500bpm Perfusion index: 0.05%- -20% * Accuracy range: SpO2 saturation: 70-100% Pulse rate: 20-500BPM Perfusion Index: 0.025%-20% |
Measurement accuracy |
* SpO2: Big animal (including pediatric ) +2digit (70-100%) not defined ( <70% ) Small animal +3 digits not defined ( <70% ) On motion condition +3digit * PR: big/small animal (including pediatric) + 3digits On motion condition + 3digit * Perfusion Index:0.05%-20% |
Resolution |
SpO2 saturation: 1% Pulse rate: 1BPM |
Temperature |
|
Measurement range |
25-45°C |
Resolution |
0.1°C |
Accuracy |
0.1°C |
CO2 (M3E) |
|
Measurement range |
ETCO2: 0-150mmHg (0-19.7%) (0 20kPa) RR: 3-150BPM |
Accuracy range |
ETCO2 Range Accuracy 0-40mmHg ±2mmHg 41-70mmHg ±5% of reading 71-100mmHg ±8% of reading 101-150mmHg ±10% of reading RR Accuracy: ±1BPM |
Resolution |
ETCO2: 0.1mmHg; RR: 1BPM |
Setting compensation parameters |
Pressure range: 450-850mmHg Oxygen range: 0-100% Equilibrium gas: N2O, He |
Power Requirements |
5.0vdc±5%, 1A(Module average power consumption is less than 1.3W) |
Temperature |
Work Environment: 10-40°C Storage Environment: -40-70°C |
Humidity |
Work Environment: 10%-90% RH Storage Environment: <90% RH |
Power Type |
Lithium ion battery |
Operation time |
8 hours for normal operation |
Temperature |
Operating: 0°C~45°C Storage Transportation: -20°C-65°C |
Humidity |
Operating: 15% to 90% non condensing Storage: 0% to 95% non condensing |
Altitude |
Operating Altitude: Up to 5000 meters |
Dimension |
146mm*67mm*30mm |
Weight |
250g |